ASTM D5293 Apparent Viscosity By Cold Cranking Simulator CCS
Brand KN
Product origin Dalian,China
Delivery time deliver as soon as received payment
Supply capacity 30 sets one month
The CCS apparent viscosity of automotive engine oils correlates with low temperature engine cranking. CCS apparent viscosity is not suitable for predicting low temperature flow to the engine oil pump and oil distribution system. Engine cranking data were measured by the Coordinating Research CouncilThe measurement of the cranking viscosity of base stocks is typically done to determine their suitability for use in engine oil formulations. A significant number of the calibration oils for this met
KN-5293 Cold Cranking Simulator (CCS) Viscosity Apparatus
Overview
KN-5293 Cold Cranking Simulator (CCS) Viscosity Apparatus conforms to the SAE J300 and ASTM D5293 Standard Test Method for Apparent Viscosity of Engine Oils and Base Stocks Between –5 and –35°C Using Cold-Cranking Simulator, which used for determining apparent viscosity of engine oils and base stocks by cold cranking simulator (CCS) at temperatures between –5 and –35°C at shear stresses of approximately 50 000 to 100 000 Pa and shear rates of approximately 105 to 104 s–1 for viscosities of approximately 900 to 25 000 mPa•s. The range of an instrument is dependent on the instrument model and software version installed. Apparent Cranking Viscosity results by this method are related to engine-cranking characteristics of engine oils.
Features
We use retro reflector to observe the experimental phenomenon
Total English version operation interface.
Omron temperature controller, Light touch button with convenient operation.
The apparatus can detect rotary speed automatically, use fine tuning knob to control current, reduce much manual operation error.
Equipped with double imported compressor as cooling system, refrigeration depth can reach -60℃.
The apparatus is equipped with imported click drive with good precision.
It can print and store test result.
The apparatus will calculate result automatically after calibrating with CANNON CL160 viscosity standards
We adopt embedded operating system with highly stable and reliable working application.
Second generation rotor, low torque testing condition with high repeatability
The apparatus will automatically shut down and start to heat after each experiment, this function is good for rapid clean.
The apparatus recommends refrigeration speed automatically.
The apparatus is equipped with digital rotary encoder to detect rotary speed automatically.
User can edit and store all standard oil’s reference value
The apparatus has a capacity of storing over 1000 tests data, very easy to check.
Software has High viscosity function to avoid Weissenberg effect
Technical Parameters
Rated Voltage:AC220V±20%
Power: 1500W
Temperature Range: external circulated alcohol bath temperature control range ~-60℃
Cooling Bath Temperature accuracy:±0.1℃
Stator Temperature Accuracy:±0.05℃
Temperature display accuracy: ±0.01℃
Viscosity Measuring Range:1500~27000mPa.s
Ambient Requirements:temperature 10~40℃;humidity≤85%
We use CANNON standard to calibrate our CCS.