ASTM D7279 Automated Houillon Viscometer
Brand KN
Product origin DALIAN
Delivery time 30 DAYS AFTER RECEIVING FULL PAYMENT
Supply capacity 30 sets/month
KN-7929 Automated Houillon Viscometer conforms to ASTM D7279 Standard Test Method for Kinematic Viscosity of Transparent and Opaque Liquids by Automated Houillon Viscometer. Many petroleum products and some non-petroleum products are used as lubricants in the equipment, and the correct operation of the equipment depends upon the appropriate viscosity of the lubricant being used. Additionally, the viscosity of many petroleum fuels is important for the estimation of optimum storage, handling, and operational conditions. Thus, the accurate determination of viscosity is essential to many product specifications.
KN-7279 Automated Houillon Viscometer
Overview
KN-7929 Automated Houillon Viscometer conforms to ASTM D7279 Standard Test Method for Kinematic Viscosity of Transparent and Opaque Liquids by Automated Houillon Viscometer. Many petroleum products and some non-petroleum products are used as lubricants in the equipment, and the correct operation of the equipment depends upon the appropriate viscosity of the lubricant being used. Additionally, the viscosity of many petroleum fuels is important for the estimation of optimum storage, handling, and operational conditions. Thus, the accurate determination of viscosity is essential to many product specifications.
The viscosity of used oils is a commonly determined parameter in the oil industry to assess the effect of engine wear on the lube oils used, as well as the degradation of the engine parts during operation.
The Houillon viscometer tube method offers automated determination of kinematic viscosity. Typically a sample volume of less than 1mL is required for the analysis.
Features
Fast analysis, be able to get the results within 15s under the fast analysis mode. The results are generally obtained within 3 minutes. The whole process of sample injection, test, cleaning, drying, and result calculation generally takes no more than 7 minutes.
High automation. Test, cleaning, drying, and result calculation are all done automatically
Be able to test transparent and opaque samples, including gasoline, diesel, kerosene, cutting fluid, heat transfer oil, additives, new oil for lubricating oil and oil in use and other Newtonian liquids
Fully automatic cleaning, fast cleaning and low cleaning cost. A cleaning process generally consumes no more than 10 ml cleaning solution
Many auxiliary functions: constant calibration, temperature calibration, internal clock timing verification, automatic calculation of viscosity index, automatic conversion of kinematic viscosity to Enn's viscosity
Software and hardware dual over-temperature protection, over-temperature alarm, anti-dry burning protection function
The heating and cooling speed is fast. The fastest heating and cooling rate is 5℃/min
Adopts double-layer glass cylinder, the temperature is more uniform
Adopts PT500 high-precision temperature sensor, the temperature of the constant temperature bath is stable and accurate, and the temperature control accuracy reaches 0.01 ℃
It can be equipped with a disposable filter, which greatly reduces the workload of the operator on sample pretreatment
Quick viscometer replacement without emptying the bath
Be able to transmit data through WIFI, connect to LIMS system
Technical parameters
Rated voltage: AC220V±10% 50Hz
Ambient temperature: 10~28℃
Relative humidity<80%
Measuring range: 0.3~6000mm²/s, the measuring span of a single viscometer is 10 times
Sample volume: 0.3~1ml
Temperature control range: 20~100℃(chiller is optional, if the expected temperature is lower than 40℃)
Temperature control accuracy: 0.01℃
Timing accuracy: 0.01s
Repeatability: ≤0.05%
Total power: ≤300W
Dimension: 330*200*450mm
There should be no strong vibration, airflow, strong electromagnetic interference and corrosive gas around the laboratory