Automatic Distillation Apparatus For Petroleum Products
Brand KN
Product origin Dalian,China
Delivery time deliver as soon as received payment
Supply capacity 30 sets one month
KN-86D Automatic Distillation Apparatus for Petroleum Products is a new designed distillation tester. It conforms to standard ASTM D86 Standard Test Method for Distillation of Petroleum Products at Atmospheric Pressure , which is suitable for testing distillation of light and middle distillates, automotive spark-ignition engine fuels with or without oxygenates, aviation gasoline, aviation turbine fuels, diesel fuels, biodiesel blends up to 20 %, marine fuels, special petroleum spirits, naphthas,
KN-86Z Automatic Distillation Apparatus for Petroleum Products
Overview
KN-86Z Automatic Distillation Apparatus for Petroleum Products is a new designed distillation tester. It conforms to standard ASTM D86 Standard Test Method for Distillation of Petroleum Products at Atmospheric Pressure , which is suitable for testing distillation of light and middle distillates, automotive spark-ignition engine fuels with or without oxygenates, aviation gasoline, aviation turbine fuels, diesel fuels, biodiesel blends up to 20 %, marine fuels, special petroleum spirits, naphtha, white spirits, kerosene, and Grades 1 and 2 burner fuels.
Features
The apparatus can automatically measure and control all experimental process by the single chip microcomputer, 10-inch touching LCD screen displays all experimental process and show the temperature data, volume data and related curves
Accept the cylinder bath temperature via automatic control
The apparatus adopts quartz base plate
Automatic nitrogen outfire is OK
Track the liquid level automatically and control the flow velocity
Technical parameters
Rated voltage:AC220V±10% 50Hz or 110V±10% 60Hz
Power:1200W
Temperature range: Environment temperature~400℃
Temperature control Precision:±0.1℃
Temperature sensor: RTD Pt100
Start heating to Initial distillation point time:5~15min
Distillation rate:4~5min
Residues in flask is 5 ml: less than 5 min
Initial distillation point temperature detection: Optical fiber sensors
Recycling volume measurement range:0~100ml
Recycling volume measurement precision:±0.1ml
Ambient requirements: Temperature 10℃~40℃;Humidity≤85%